Full details for "Complement Function". | |
---|---|
Name |
Complement Function |
Alternative name/Profile |
CH100, CH50, Alternative Pathway, Classical Pathway, Mannose-binding Lectin Pathway (MBL), Haemolytic Cascade, C1, C1q, C1r, C1s, C2, C3, C4, C5, C6, C7, C8, C9, Membrane Attack Complex (MAC), Factor B, Factor D, Factor H, Factor I, Properdin, MBL-Associated Serine Proteases (MASPS) |
Department |
Immunology |
Investigation |
Enzyme Immunoassay (ELISA) (Euro Diagnositca, Weislab Complement System) |
Specimen type |
Clotted blood |
Sample type |
Serum |
Sample container & volume |
5ml Red |
Frequency of analysis |
Batch basis |
Turnaround time |
21 days |
Availability |
|
Notes |
Useful for:
Clinical Information:
Please Note:
|
Related links |
|
Reference range |
Normal Complement Cascade Activation: Classical Pathway: 32.4-125% Alternative Pathway: 24.5-113% Reduced Complement Cascade Activation: (Repeat sample advised for verification) Classical Pathway: 5-32.4% Alternative Pathway: 5-24.5% Reference ranges determined with 13 normal patient samples, and confirmed as normal using radial immunodiffusion assays in SJH 2014. Absent Complement Cascade Activation : Classical and Alternative Pathway: <5% Alternative-Absent, Total-Normal: Indicates a deficiency in the Alternative Pathway. Repeat sample advised for verification. Alternative-Absent, Total-Absent: Indicates a deficiency in the Terminal Pathway. Repeat sample advised for verification. Alternative-Normal, Total-Absent: Indicaties a deficiency in the Classical Pathway. Repeat sample advised for verification. |
Last updated |
Wed, 15 Mar 2023 12:02:55 GMT |